9-20
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
General Maintenance: Maternal Blood Pressure Tests
Patient-to-Ground Leakage for FSpO
2
1.
Connect an FSpO
2
test body to the monitor’s front panel
FSpO
2
input. Use the
assembly leakage cable for the test body.
2.
Set the fixture’s
Leakage
switch to
Patient
.
3.
Turn ON the monitor using the monitor’s power switch.
4.
Verify the following:
F
< 50
µ
A (100, 120, 220, 230, or 240 VAC monitors)
5.
Record the patient-to-ground leakage for the following conditions:
Table 9-9. Patient-to-Ground Leakage Conditions for FSpO
2
Neutral
Ground
Power (Polarity)
Closed
Closed
Normal
Closed
Open
Normal
Closed
Open
Reversed
Closed
Closed
Reversed
Open
Closed
Normal
Open
Closed
Reversed
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......