Self-Tests: Error Log Screen
Error Log Screen
The Error Log screen,
Figure 8-2
, displays a service log of the 120 Series Monitor.
The service log is helpful to service representatives when diagnosing intermittent
problems. There are up to 230 error codes which can be detected by the monitor.
The Error Log is comprised of multiple pages with up to 20 error codes per screen.
Each screen displays three columns:
The Error ID column lists the error code. The codes are defined in
Table 8-1
.
The Count column lists the number times the error code has occurred.
The Data column lists other information associated with the error code. This
column is for factory use only.
Figure 8-2. Error Log Screen
PRINT
CLEAR
PAGE
EXIT
ERROR LOG
ERROR LOG
COUNT
DATA
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......