Revision B
120 Series Maternal/Fetal Monitor
6-7
2015590-001
Functional Checkout Procedure: MECG Test
MECG Test
This portion of the functional checkout procedure ensures the integrity of the MECG
circuitry and the heart rate channel of the recorder.
1.
Connect the simulator’s ECG sub-cable to the
MECG
connector on the monitor
adapter cable.
2.
Connect the simulator’s UA sub-cable to the
UA
receptacle on the monitor.
3.
Set the switches on the Model 325 Input Simulator according to
Table 6-1
.
4.
If not already on, depress the monitor’s
Record
pushbutton.
5.
Turn the simulator’s
Manual Adjustment
knob fully counterclockwise and
verify the following on the monitor’s display:
The MHR value is 30 BPM.
The MHR/P mode is
MECG
.
The MHR heartbeat indicator (
) flashes at a rate of 30 times per minute.
The UA mode is
TOCO
.
6.
Access the Install Options screen and note the default TOCO reference value.
7.
Press the monitor’s front panel
UA Reference
pushbutton.
8.
Verify the following on the monitor:
The UA value is referenced to the default value.
The recorder prints a continuous line at the default value on the bottom grid
of the strip chart paper.
The recorder prints the messages
MECG
,
TOCO
, and
UA REF
on the strip
chart paper.
9.
Turn the simulator’s
Manual Adjustment
knob to input an MECG signal of
approximately 60 BPM. Verify the following on the monitor:
The MHR value is 60 BPM.
The MHR heartbeat indicator (
) flashes at a rate of 60 times per minute.
The ECG “beep” volume can be heard from the rear panel speaker. The
volume can be adjusted on the MHR/P Setup screen.
The recorder prints a continuous line at 60 BPM on the top grid of the strip
chart paper.
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......