10-14
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Strip Chart Recorder Servicing: Field Serviceable Assemblies
Paper-Loading Sensor
Follow the instructions for
“Removing the Strip Chart Recorder”
on
page 10-2
before following this procedure:
Removing the Paper-Loading Photosensor
1.
Disconnect the paper-loading sensor cable from J8 on the Recorder Board.
2.
Turn over the recorder and place face down.
3.
Cut and remove the cable tie which holds the sensor cable.
4.
Remove the two screws which hold the sensor and housing to the paper tray.
5.
Discard the sensor and printed circuit board.
Replacing the Paper-Loading Sensor
Use the following procedure for replacing the paper-loading photosensor:
1.
Secure the sensor and printed circuit board to the paper tray and tighten both
screws.
2.
Connect the sensor cable to J8 on the Recorder Board.
3.
Install a new cable tie to hold the cable in place.
NOTE:
Ensure that there is enough slack in the cable so that when the recorder
door is opened the cable does not become taut.
4.
Re-install the strip chart recorder. Refer to
“Installing the Strip Chart Recorder”
on
page 10-2
. Do not replace the monitor top cover yet.
5.
Adjust the paper-loading sensor as described next in this section.
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......