Revision B
120 Series Maternal/Fetal Monitor
9-13
2015590-001
General Maintenance: Maternal Blood Pressure Tests
Unit to Primary Leakage
1.
Set the fixture’s
Leakage
switch to
Unit
.
2.
Turn ON the monitor using the monitor’s power switch.
3.
Verify the following:
F
< 300
µ
A (100, 120, 220, 230, or 240 VAC monitors)
4.
Record the primary leakage for the following conditions:
Table 9-2. Unit to Primary Leakage Conditions
Neutral
Ground
Power (Polarity)
Closed
Closed
Normal
Closed
Open
Normal
Closed
Open
Reversed
Closed
Closed
Reversed
Open
Closed
Normal
Open
Open
Normal
Open
Open
________ Reversed
Open
Closed
Reversed