6-22
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Functional Checkout Procedure: Uterine Activity Test
Uterine Activity Test
This portion of the functional checkout procedure tests the uterine activity section of
the 120 Series Monitor.
1.
Set the switches on the Model 325 Simulator according to
Table 6-5
.
2.
Connect the simulator’s UA sub-cable to the
UA
receptacle on the monitor.
3.
Access the Install Options service mode screen and note the default TOCO
reference value. (The monitor is shipped from the factory with this value set at
10 relative units; however, your unit may have been custom configured.) Exit
the service mode screens.
4.
If not already on, depress the monitor’s
Record
pushbutton.
5.
Briefly press the monitor’s
UA Reference
pushbutton. Verify the following on
the monitor:
The UA value is the
default
setting.
The UA mode is
TOCO
.
The recorder prints a continuous line at the
default
value on the uterine
activity channel of the strip chart paper.
The recorder prints the messages
TOCO
and
UA REF
on the strip chart
paper.
6.
Press and hold the
UA Reference
pushbutton on the monitor to cycle through
the available selections for UA reference: 5, 10, 15, 20, or 25 relative units.
Test each of these reference settings. Verify that the UA value is displayed
accordingly and that the recorder prints a continuous line at the corresponding
value on the uterine activity channel of the strip chart paper.
7.
Place the simulator’s
UA Level
switch at each of the level settings: 0, 10, 50,
and 100 relative units. Verify that the UA value is displayed accordingly and
that the recorder prints a continuous line at the corresponding value on the heart
rate channel of the strip chart paper.
8.
Place the simulator’s
UA Mode
switch in the IUP position and the
UA Level
switch to 0 mmHg. Depress the monitor’s
UA Reference
pushbutton and verify
that the monitor and recorder reference to 0 mmHg. Verify the following on the
monitor:
The UA value is 0 mmHg.
The UA mode is
IUP
.
The recorder prints a continuous line at 0 mmHg on the uterine activity
channel of the strip chart paper.
The recorder prints the messages
IUP
and
UA REF
on the strip chart paper.
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......