B-2
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Alarms Summary: Table of Alarms
Table of Alarms
Table B-1. Summary of 120 Series Alarms
Type
Condition
Visible Advisory
Audible Advisory
Paper Speed
United States: 3 cm/min
International: 1 cm/min
1-3 cm/min
FHR
An alarm setting (audio or high/low
limit) is turned off.
Alarm Defaults
Audio: on
Volume: 5
Limits: High = 160 BPM, Low = 120
BPM
displays to the left of the
FHR mode title.
—
FHR limit (high or low) actively being
violated.
or
Unsilenced, resolved FHR limit
violation (the limit was violated but
the FHR has since returned to the
normal range before clinical
acknowledgement).
For continuous limit violations: a high
alarm activates after 5 minutes; a
low alarm activates after 30
seconds.
About Latching Alarms: The FHR
limit alarms are latching alarms
which means that a clinician must
acknowledge the alarm using the
monitor’s
Alarm Silence
button in
order to clear the alarm.
FHR numeric flashes.
Alternating high/low tones (if
audio enabled).
Inadequate FHR signal quality.
Flashing dashes “– – –” in place
of FHR numeric.
Alternating high/low tones (if
audio enabled).
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......