Revision B
120 Series Maternal/Fetal Monitor
6-35
2015590-001
Functional Checkout Procedure: MSpO
2
Test
MSpO
2
Test
1.
Access the 120 Monitor’s MSpO
2
Setup window and configure as follows:
(Nellcor only) Response Time = Normal
(Masimo only) Sensitivity = Normal
(Masimo only) Averaging = 8
Print Interval = 2 minutes
Trace = On
2.
Access the monitor’s MSpO
2
Setup window and set the MSpO
2
Scale to AUTO.
3.
Access the monitor’s MHR/P Setup window and configure as follows:
MHR/P Source = MSpO
2
MHR/P Trace = On
4.
If your monitor has Nellcor technology installed, connect one end of the Nellcor
cable to a Nellcor pulse oximetry simulator; connect the other end to the
monitor’s MSpO
2
connector.
5.
If your monitor has Masimo technology installed, connect one end of the
Masimo cable to a Masimo pulse oximetry simulator; connect the other end to
the monitor’s MSpO
2
connector.
6.
Allow the monitor a few seconds to obtain a steady reading.
7.
Turn on the recorder and allow data to collect for at least five minutes.Verify
the following on the monitor:
The correct waveform appears on the display.
The %MSpO
2
display value is 81% ± 2%
The MSpO
2
pulse amplitude indicator shows a fluctuating bar graph.
The MHR/P display mode is PULSE.
The MHR/P display value is 112 BPM ± 3 BPM
The MHR/P trend plots in the top grid at 112 BPM ± 3 BPM.
The MSpO
2
scale grid marks stamp on the paper.
The message MSpO
2
% stamps in the annotation line on the strip chart
paper.
A diamond
(with %MSpO
2
and MHR/P vital signs) stamps in the
annotation area on the strip chart paper at two-minute intervals.
NBP Test
Refer to
“Accessing the Diagnostic Control Screen”
on page 9-6 for information
about a manometer pressure check and pneumatic pressure check.
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......