Revision B
120 Series Maternal/Fetal Monitor
7-3
2015590-001
Calibration: Power Supply Voltages—Verification
Power Supply Voltages—Verification
There are no adjustments for the power supply voltages; however, you can verify
them as follows:
Main Board Power Supply Voltages
Using a digital voltmeter, measure and verify the power supply voltages using the
J14 connector on the Main Board. Refer to
Table 7-1
.
NOTE:
There are no adjustments for these voltages.
Table 7-1. Main Board Power Supply Voltages
J14 ON 11619
Signal Name
Voltage Level
Pin 1
+12EL
+12 Vdc ± 0.5 Vdc
Pin 2
+20I
+20 Vdc ± 0.5 Vdc
Pin 3
+15BP
+15 Vdc ± 0.5 Vdc
Pin 4
–15V
–15 Vdc ± 0.5 Vdc
Pin 5
+15V
+15 Vdc ± 0.5 Vdc
Pin 6
+12A
+12 Vdc ± 0.5 Vdc
Pin 7
+5V
+5 Vdc ± 0.5 Vdc
Pin 8
GND
—
Pin 9
No Connection
—
Pin 10
Keying
—
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......