15-4
120 Series Maternal/Fetal Monitor
Revision B
20155
90
-001
Troubleshooting: FECG Troubleshooting
FECG Troubleshooting
Table 15-3. FECG Troubleshooting
Problem
Probable Cause
Possible Solution
Internal FECG erratic or not recording
properly.
Transducer not properly connected to
monitor.
Attachment pad or legplate not securely
attached to patient.
Electrode wire not secure in
legplate post.
Paste is dried or incorrect paste is being
used.
Electrode not properly attached.
No FECG signal.
Defective electrode.
Defective attachment pad.
Service required.
Ensure transducer is securely attached
to the monitor.
Secure attachment pad or legplate to
patient.
Inspect legplate connection.
Check ECG paste; re-apply, if
necessary.
Replace electrode.
Auscultate FHR.
Replace electrode.
Replace attachment pad.
Call Biomedical Engineering
Department.
Rate in FHR area of the display and the
FHR trend on the strip chart paper do not
correlate.
Monitor is set for 30 BPM/cm vertical scale
and 20 BPM/cm vertical scale strip chart
recorder paper is being used (or vice
versa).
Call Biomedical Engineering Department.
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......