Revision B
120 Series Maternal/Fetal Monitor
9-17
2015590-001
General Maintenance: Maternal Blood Pressure Tests
Patient-to-Line Leakage for IUP
1.
Connect an IUP test body to the monitor’s front panel
UA
input. The IUP test
body can be a SensorTip transducer (wrapped in aluminum foil) or the assembly
leakage cable.
2.
Set the fixture’s
Leakage
switch to
Line
.
3.
Turn ON the monitor using the monitor’s power switch.
4.
Verify the following:
F
< 50
µ
A (100, 120, 220, 230, or 240 VAC monitors)
5.
Record the patient-to-ground leakage for the following conditions:
Table 9-6. Patient-to-Line Leakage Conditions for IUP
Neutral
Ground
Power (Polarity)
Closed
Closed
Normal
Closed
Closed
Reversed
Open
Closed
Normal
Open
Closed
Reversed