Revision B
120 Series Maternal/Fetal Monitor
6-37
2015590-001
Functional Checkout Procedure: Preventative Maintenance Inspection Report
Inspection Checklist
Page
Description
Complete
Max
Actual
Pass
Fail
N/A
Cleaning
page
9-2
Clean the monitor exterior.
Clean the monitor interior using a hand-held
vacuum.
page
10-4
Clean the recorder printhead.
page
9-3
Clean the monitor accessories (tocos,
transducers, cables)
page
9-4
Clean the monitor accessories (cuffs, hoses)
Calibration
Page
Description
Complete
Max
Actual
Pass
Fail
N/A
page
7-3
Verify the power supply voltages on the Main
Board.
page
7-4
Verify the power supply voltages on the
Isolated Power Supply Board.
page
10-6
Adjust the printhead.
page
10-12
Adjust the photosensors on the Recorder Board.
Electrical Safety Tests
Page
Description
Complete
Max
Actual
Pass
Fail
N/A
page
9-12
AC Line (120 V
10%)
(<300 m
Ω)
page
9-12
Ground Impedance (<300 m
Ω)
Lead to rear panel ground lug (<300 m
Ω)
Lead to top cover bezel screw (<300 m
Ω)
(<300 m
Ω)
page
9-13
Unit to primary leakage
(<300
µ
A)
page
9-14
Patient-to-Ground leakage for ECG
(
<
50µ
A)
page
9-15
Patient-to-Line leakage for ECG
(
<
50µ
A)
page
9-16
Patient-to-Ground leakage for IUP
(
<
50µ
A)
page
9-17
Patient-to-line leakage for IUP
(
<
50µ
A)
↑
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......