B-4
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Alarms Summary: Table of Alarms
NOTE:
You must cycle power to reset factory defaults.
MSpO
2
b
An alarm setting (audio or high/low
limit) is turned off.
Alarm Defaults
Audio: on
Volume: 5
Limits: High = 100%, Low = 95%
Re-alarm: 120 sec
displays to the left of the
MSpO
2
mode title.
—
%MSpO
2
limit (high or low) actively
being violated. Issued after about 8
seconds.
%MSpO
2
numeric flashes.
%MSpO
2
value and pulse rate
print on the strip chart.
Alternating high/low tones (if
audio enabled).
Malfunction with MSpO
2
circuitry.
COMM
or
REPAIR
message
displays in MSpO
2
area.
Alternating high/low tones (if
audio enabled).
MSpO
2
intermediate cable
disconnected from monitor; sensor
assembly disconnected from
intermediate cable; or sensor or
cable has a broken wire.
Dashes “– – –” in place of
%MSpO
2
numeric.
Alternating high/low tones (if
audio enabled).
a
There is an MECG re-alarm.
b
There is an MSpO
2
re-alarm.
Table B-1. Summary of 120 Series Alarms
Type
Condition
Visible Advisory
Audible Advisory
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......