Revision B
120 Series Maternal/Fetal Monitor
3-3
2015590-001
Controls, Indicators, and Connectors: Front Panel Description
Table 3-1. Model 129 Front Panel
#
Name
Description
A
Display
The monitor’s display is divided into several sections. The content and layout of the
display can change, depending on which functions are installed in the monitor and the
modes of operation in use.
B
FHR1 Volume Decrease Button
The four Volume buttons raise (
) and lower (
) the volume of sound emitted by
the rear panel speaker. The left pair controls the volume for FHR1. The right pair
controls the volume for FHR2.
Volume settings have no effect on the processing used to determine heart rate. The
volume buttons work in conjunction with the volume control settings on the US/US2
Setup screen (
page 5-9
) and on the FECG Setup screen (
page 5-8
).
C
FHR1 Volume Increase Button
D
FHR2 Volume Decrease Button
E
FHR2 Volume Increase Button
F
Alarm Silence Button
Pressing this button removes the audible indication of an individual alarm.
G
Test Button
Pressing and holding this button for one second starts or stops a monitor self-test
routine.
H
BP Start/Stop Button
This button starts and stops both manual and automatic blood pressure
determinations. It also provides a “shortcut” for changing the auto interval time.
I
UA Reference Button
The UA Reference button sets a baseline for uterine activity pressure monitoring.
J
Mark [Offset] Button
The Mark (Offset) pushbutton is a multi-function pushbutton.
Mark: Pressing this button prints an event mark on strip chart paper (on the
bottom two lines of the top grid).
Offset: When the Heart Rate Offset mode is enabled, pressing and holding this
button shifts the secondary FHR trend +20 BPM for visibility purposes.
K
Paper Advance Button
Pressing this button advances chart paper at a rate of 40 cm/min for as long as the
button is held down.
L
Record Button
The Record button selects one of three recorder states: on, maternal-only mode, or off.
M
Power Indicator
The indicator lights green when the monitor is turned on.
N
Record Indicator
Indicator Status
Recorder Status
on
on
off
off
flickers
maternal-only mode
flashing
error condition
O
Trim Knob Control
Operation of the monitor is controlled by using the front panel pushbuttons in
conjunction with the Trim Knob control. This control selects softkeys on the display
and positions a cursor within a setup screen. Rotate the Trim Knob control left or right
to highlight items on the screen with a bar cursor. After highlighting the desired item,
press the Trim Knob control to make the selection. In summary: rotate to move cursor;
press to select an item.
P
Recorder Door Latch
Opens the strip chart recorder door to add, remove, or adjust the paper.
Q
Power Switch
Moving the switch to the on position (
I
) turns on the monitor; moving the switch to the
off position (
O
) turns off the monitor.
R
Strip Chart Recorder
Annotations and trends are printed on the strip chart paper. Two paper styles are
available. Refer to
“Chapter 5, Setup Procedures”
, for instructions on loading strip
chart paper into the recorder.
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......