Revision B
120 Series Maternal/Fetal Monitor
8-15
2015590-001
Self-Tests: Diagnostic Control Screen
NO ADDRESS
Wallplate cable may be disconnected; or Corolan address checking is
disabled.
NO DATA YET
Contact GE Service Representative.
NEVER INTERRUPTED
There is no communication with the Corolan master device (e.g. QS Server).
The master device is not polling the monitor.
NO RECENT INTERRUPT
The monitor was successfully communicating with the Corolan master
device (e.g. QS Server); however, two minutes have passed since activity.
DRIVER NOT RUN
Contact GE Service Representative.
TRANSMITTING
LONG TRANSMISSION
IN TEST MODE
NO RECENT RECEPTION
PACKET ERROR
PACKET TOO SHORT
SIM RECEIVED
DISC RECEIVED
OFFLINE ADDRESS CONTENTION
Two devices (monitors) have the same wallplate address. Both devices
have automatically been taken offline.
WAS OFFLINE
Contact GE Service Representative.
HAD DISC ADDRESS
CRC ERROR
UNKNOWN PACKET TYPE
RECENT ADDRESS CONTENTION
A monitor has just been turned on using the same wallplate address as
another monitor. If the situation continues, this message will change to
OFFLINE ADDRESS CONTENTION. Disconnect both monitors from
wallplate. Check wallplate addresses. Re-connect.
SOME ADDRESS CONTENTION
PACKET NOT A POLL
Contact Corometrics Service Representative.
NO NETWORK ACTIVITY
Twenty seconds have elapsed since last network activity detected.
15 SEC SINCE ACTIVITY
Fifteen seconds have elapsed since last network activity detected.
10 SEC SINCE ACTIVITY
Ten seconds have elapsed since last network activity detected.
5 SEC SINCE ACTIVITY
Five seconds have elapsed since last network activity detected.
AWAITING MANAGER DISCONNECT
Contact Corometrics Service Representative.
NO SNRM YET
Table 8-3. Corolan Network Status Messages
Message
Meaning
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......