10-6
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Strip Chart Recorder Servicing: Field Serviceable Assemblies
Printhead Adjustments
Refer to
Figure 10-1
and
Figure 10-2
for identification of the adjustment points
outlined in this section, and to assist in clarifying adjustment procedures.
Figure 10-
1
and
Figure 10-2
show top and right side views of the 120 Series Monitor strip
chart recorder assembly removed from the unit’s chassis.
Table 10-1
gives a description of each recorder adjustment. Adjustments C through
F can be performed without removing recorder from unit’s chassis; however, they
require removing the monitor’s top cover.
The printhead may require performing one or both of the following offset
adjustment procedures.
Vertical Offset Adjustment
To check vertical offset alignment, run at least two monitor self-tests by depressing
the front panel
Test
pushbutton. The first horizontal trace should be printed 0.490 ±
0.002 inch from the right hand paper guide. (Refer to
Figure 10-3
.) Perform the
following steps if the trace does not fall within the specified offset range.
1.
Loosen the two set screws labeled
A
in
Figure 10-1
.
2.
Adjust hex-head screw labeled
C
in
Figure 10-2
, until test trace falls within
specified tolerance.
3.
Tighten the two set screws labeled
A
.
Horizontal Offset Adjustment
Adjust the horizontal offset if any of the following symptoms are detected. If
skewing occurs, check for other malfunctions. Noticeable skewing of a vertical line
printed on the strip chart paper is usually accompanied by one of the following
symptoms:
Paper consistently curls to one side.
Printing of unequal weight occurs along the vertical trace line, from one end to
the other.
Dots are missing along the vertical trace.
Printing is too light following printhead replacement.
If the Horizontal Offset Adjustment is required, perform the following steps:
1.
Loosen the four hex-head lock screws—two on each side. The right side screws
are labeled
B
in
Figure 10-2
.
2.
To move the printhead forward on one side, back-off the corresponding captive
screw (turn counterclockwise) from its alignment block. To move the printhead
backward on one side, tighten the corresponding captive screw (turn
clockwise).
3.
After making the necessary adjustments, tighten the four hex-head lock screws
labeled
B
.
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......