Revision B
120 Series Maternal/Fetal Monitor
10-3
2015590-001
Strip Chart Recorder Servicing: Installing the Strip Chart Recorder
Installing the Strip Chart Recorder
1.
Carefully place the recorder into the monitor housing.
2.
Install the four front recorder mounting screws which secure the frame to the
bezel.
3.
Re-connect the Power Supply cable harness to the Recorder Board at J1.
4.
Re-connect the Main Motherboard ribbon cable to the Recorder Board at J2.
5.
Re-connect the DSP ribbon cable to the membrane switch panel.
6.
Re-connect the display ribbon cable to the DSP Board at J5.
7.
Carefully reposition the front bezel. Secure all four screws.
8.
Install the two screws which secure the recorder to the bottom panel.
9.
Replace the copper grounding strips at the top of the front bezel. Be careful not
to damage any of the fingers.
10. Replace the monitor top cover and secure with all nine screws.
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......