Revision B
120 Series Maternal/Fetal Monitor
6-17
2015590-001
Functional Checkout Procedure: Ultrasound Test
8.
Place the simulator’s
US Rate
switch in each of the individual rate settings (50,
60, 120, and 210 BPM). Verify the following on the monitor:
The FHR1 value reflects the simulator setting ± 1 BPM.
The FHR1 heartbeat indicator (
) flashes at the simulator setting.
Ultrasound audio is heard coming from the rear panel speaker.
The recorder prints a continuous line at the respective value ± 3 BPM on
the top grid of the strip chart paper.
9.
Repeat
step 4
through
step 8
using the second ultrasound channel. (The mode
will show
US2
.)
10. Place the simulator’s
US Mode
switch in the OFF position. Verify the
following on the monitor:
The FHR1 value and mode are both blank.
The recorder stops printing the fetal heart rate trace.
The recorder prints the message
CARDIO INOP
on the center margin of the
strip chart paper after approximately 20 seconds.
Table 6-3. Ultrasound Test Simulator Settings
Section
Switch
Setting
US/FMD
Mode
US
Signal Level
MED
Rate
MANUAL
GENERAL
Pattern Memory
OFF
UA
Main
CMR
Mode
TOCO
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......