Revision B
120 Series Maternal/Fetal Monitor
10-7
2015590-001
Strip Chart Recorder Servicing: Field Serviceable Assemblies
Table 10-1. Printhead Adjustments
Adjustment
Description
A
Set screws loosened to perform a vertical
offset adjustment.
B
Hex-head screw loosened to perform a
horizontal offset adjustment.
C
Hex-head screw which determines the
vertical offset of a horizontal line. (See
Figure 10-3
.)
D
Captive hex-head screw which controls
the horizontal offset of a vertical line on
the top grid of the strip chart paper (heart
rate channel).
E
Captive hex-head screw which controls
the horizontal offset of a vertical line on
the bottom grid of the strip chart paper
(uterine pressure channel).