Revision B
120 Series Maternal/Fetal Monitor
15-9
20155
90-
001
Troubleshooting: Internal UA Troubleshooting
Internal UA Troubleshooting
Table 15-6. Internal UA Troubleshooting
Problem
Probable Cause
Possible Solution
Internal pressure not measuring correctly.
Transducer not properly connected to
monitor.
Air bubble in dome; or catheter blocked.
Dome is cracked.
Strain gauge not at same height as
catheter tip.
Catheter has fallen out of place.
Catheter or strain gauge not zeroed.
Service required.
Ensure transducer is securely attached
to monitor.
Flush dome and catheter.
Replace dome.
Adjust strain gauge height.
Replace catheter.
Calibrate catheter or strain gauge.
Call Biomedical Engineering
Department.
CHECK IUP
message displayed in UA area
of the display.
Blockage in fluid-filled catheter.
Fetus pressing directly on catheter.
Defective strain gauge or catheter.
Service required.
Flush catheter. Re-zero. Replace
catheter if necessary.
Reposition by twisting catheter.
Replace strain gauge or catheter.
Call Biomedical Engineering
Department.
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......