A-2
120 Series Maternal/Fetal Monitor
Revision B
20155
90
-001
Factory Defaults: Table of Defaults
Table of Defaults
Table A-1. Summary of Factory Defaults
Setup
Screen
Field Description
Factory Default
Default Options
Hospital Preference
FECG or
US/US2
HR Offset
(Applies to US or US2—
whichever is FHR2)
Off
On, Off, 10 minutes
FHR Volume
5
0-9
FHR Alarm Limits
High Low
160 120 BPM
200-140, Off
60-140, Off
Audio Alarms
On
On, Off
Volume
5
1-9
FSp0
2
Response Time (Averaging
Mode)
Slow
Slow, Fast
Print Interval
5 min
Off, 2, 5, 10, 15, 30, 60 min
% 0
2 Trace
On
On, Off
NBP
Display
On
On, 1, 2, 3, 5, 10, 15,
30 min
Mode
Manual
Manual, Auto, 1, 2, 3, 4, 5,
10, 15, 20, 30, 40, 45, 60,
90, 120 min
NBP Reading Done
5
0-9
Alarm
Systolic
Diastolic
MAP
MHR/P
High
Low
160
90 mmHg
90
50 mmHg
140
50 mmHg
120
50 BPM
High
Low
70-240, Off
50-150, Off
70-130, Off
30-120, Off
70-150, Off
30-120, Off
100-250, Off
35-120, Off
Audio Alarm
On
On, Off
Volume
5
1-9
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......