Revision B
120 Series Maternal/Fetal Monitor
8-11
2015590-001
Self-Tests: Diagnostic Control Screen
Diagnostic Control Screen
The Diagnostic Control screen,
Figure 8-3
, is used to perform the following
diagnostic self-tests:
manometer pressure check;
pneumatic pressure check;
recorder calibration test;
CPU software version number;
DSP software version number;
monitor run time;
recorder run time;
Main Board SW1 switch setting summary;
Corolan network status;
Corolan address check;
softkey to access J102 test screen; and
softkey to perform a memory dump.
Figure 8-3. Diagnostic Control
To access the Diagnostic Control screen:
1.
Select the
SERVICE
softkey from the General Setup screen.
2.
The Service Lock screen appears.
3.
Enter the correct access code; the Install Options screen displays.
4.
Select the
TESTS
softkey from the Install Options screen.
5.
The Diagnostic Control screen displays.
J102
MEM DUMP
EXIT
DIAGNOSTIC CONTROL
NBP MODE:
PATIENT
RECORDER CALIBRATION: OFF
STATUS RELAY: OFF
CPU
V3.00
DSP
03.07
RUN TIME:
72:41:58
REC TIME:
46:17:24
SW1:
1 1 1 1 0 0 0 0
MECG
MSPO2 NBP FSPO2
CORO
COROLAN: OK
COROLAN ADDRESS: 18
CLEAR
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......