A-4
120 Series Maternal/Fetal Monitor
Revision B
20155
90
-001
Factory Defaults: Table of Defaults
Master
AlarmSetup
Audio Alarms
NBP
MHR/P
MSpO
2
On
On
On
On, Off
On, Off
On, Off
Re-Alarm
(MECG and MSpO
2
only)
120 sec
30-120
(in 5 second increments)
Alarm Limits
Systolic
Diastolic
MAP
MHR/P
MSpO
2
High Low
160
90mmHg
90
50 mmHg
140
50 mmHg
120
50 BPM
100
95%
High Low
70-240, Off
50-150, Off
70-130, Off
30-120, Off
70-150, Off
30-120, Off
100-250, Off 35-120, Off
85-100, Off
80-99, Off
Volume
5
1-9
General
Time/Date
Set to current local date
and time.
Set to current local date and
time.
Play Song
Off
Off, Happy Birthday,
Brahm’s Lullaby, Rock-a-
Bye-Baby, All.
Volume
5
0-9
SpO
2
Scale
0–100 %
Auto, 0-100%
(Does not change)
Paper Speed
United States: 3 cm/min
International: 1 cm/min
1-3 cm/min
Paper Chime
Out Only
Off, Low/out, Out only
Volume
5
1-9
Light (on recorder)
On
On, Off
MSpO
2
Print Interval
(External Monitor)
5 min
Off, 2, 5, 10, 15, 30, 60 min
FSpO
2
Print Interval
(External Monitor)
5 min
Off, 2, 5, 10, 15, 30, 60 min
FSpO
2
Trace
(External Monitor)
Off
On, Off
Vital Signs
History
a
HX Interval
Event
1, 5, 10, 15, 30, 60, Event
Table A-1. Summary of Factory Defaults
Setup
Screen
Field Description
Factory Default
Default Options
Hospital Preference
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......