Revision B
120 Series Maternal/Fetal Monitor
6-23
2015590-001
Functional Checkout Procedure: Uterine Activity Test
9.
Place the simulator’s
UA Level
switch at each of the level settings: 0, 10, 50,
and 100 mmHg. Verify that the UA value is displayed accordingly and that the
recorder prints a continuous line at the corresponding value on the uterine
activity channel of the strip chart paper.
10. Place the simulator’s
UA Level
switch to the RAMP position. Verify that the
UA value measures between approximately 0 and 100 mmHg and that the
recorder prints a ramp between the same values. Refer to
Figure 6-8
.
11. Disconnect the Model 325 simulator’s uterine activity sub-cable from the
UA
input receptacle on the front panel of the monitor. Verify the following on the
monitor:
The UA value and mode are both blank.
The recorder stops printing the uterine activity trace.
The recorder prints the message
UA INOP
on the center margin of the strip
chart paper after approximately 20 seconds.
Table 6-5. Uterine Activity Test Simulator Settings
Section
Switch
Setting
GENERAL
Pattern Memory
OFF
UA
Main
LEVEL
Level
0 mmHg
Mode
TOCO
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......