Revision B
120 Series Maternal/Fetal Monitor
A-3
20155
90
-001
Factory Defaults: Table of Defaults
MSpO
2
Response Time (Nellcor)
Normal
Slow, Normal, Fast
Averaging Time (Masimo)
8 seconds
2, 4, 8, 10, 12, 14,
16 seconds
Sensitivity (Masimo)
Normal
Normal, Maximum
Print Interval
5 minutes
Off, 2, 5, 10, 15, 30,
60 minutes
%O
2
Trace
Off
On, Off
Alarms
MSpO
2
MHR/P
High
Low
100 95%
120 50
BPM
High
Low
85-100, Off
80-99, Off
100-250, Off 35-120, Off
Audio Alarms
On
On, Off
Volume
5
1-9
MHR/P
Source
Auto
Auto, MECG, MSp0
2
, NBP
MHR/P Trace
Off
On, Off
Volume
5
0-9
Alarms
High Low
120 50
BPM
High Low
100-250
35-120
Audio Alarms
On
On, Off
Alarm Volume
5
1-9
MECG Lead
II
I, II, III
Pacer Off
On,
Off
Normal
Operation
(Waveform Display)
MECG
FECG, MECG, FSp0
2
,
MSp0
2
, Off
(MECG Waveform Size)
1X (1 mV/cm)
.25X, .5X, 1X, 2X, 4X, Auto
Table A-1. Summary of Factory Defaults
Setup
Screen
Field Description
Factory Default
Default Options
Hospital Preference
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......