13-2
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Technical Specifications: General Monitor
General Monitor
Table 13-1. General Monitor Technical Specifications
Category
Technical Specifications
Power Requirements
Nominal Line Voltage:
Line Frequency:
Power Consumption (maximum):
Chassis Leakage:
100VAC
120 VAC
220 VAC
230 VAC
240 VAC
50/60 Hz
50/60 Hz
50/60 Hz
50/60 Hz
50/60 Hz
100 W
100 W
0.4 A
0.4 A
0.4 A
<300 µA
Physical Characteristics
Height:
Width:
Depth:
Weight:
6.7 in (17.0 cm)
16.5 in (41.9 cm)
17.3 in (43.9 cm)
24.0 lbs (10.9 kg) approx.
Environmental Conditions
Monitor:
Ambient Temperature:
Relative Humidity:
Atmospheric Pressure:
Strip Chart Paper
a
:
Ambient Temperature:
Relative Humidity:
Atmospheric Pressure:
a Paper operating environmental conditions are for a period of less than one month. Paper storage environmental conditions are for extended storage.
Operating
Storage
50°F to 104°F (10°C to 40°C)
14°F to 131°F (–10°C to 55°C)
10% to 95%, non-condensing
0% to 95%, non-condensing
700–1060 mbar (525–795 mmHg)
700–1060 mbar (525–795 mmHg)
50°F to 104°F (10°C to 40°C)
< 80°F (< 26.5°C)
30% to 70%, non-condensing
45% to 65%, non-condensing
700–1060 mbar (525–795 mmHg)
700–1060 mbar (525–795 mmHg)
Certification
ANSI/AAMI EC13-1992:
UL-2601.1:
CUL:
Complies with all areas except those listed below:
3.1.2.1e: Heart Rate Meter Accuracy and Response to Irregular Rhythm (not tested)
3.2.6.1: Range of QRS wave amplitude and duration
3.2.7: Range and accuracy of heart rate meter (4.2.7f :input rate of 300 BPM.)
3.2.8.1: Lower Alarm Limit (The lowest alarm limit on the 120 Series is 35 BPM.)
3.2.9.8c: Impulse Response
4.2.9.7 Output Display a) Channel Width
Classified to UL-2601.1
Medical electrical equipment classified by Underwriter’s Laboratories, Inc., with respect to
fire, shock, and mechanical hazards in accordance with UL-2601.1.
Classified with respect to electric shock, fire, mechanical, and other specified hazards
only, in accordance with CAN/CSA C22.2 No. 601.1
!
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......