Troubleshooting: General Troubleshooting
General Troubleshooting
Table 15-1. General Troubleshooting
Problem
Probable Cause
Possible Solution
No monitoring functions and green Power
indicator does not illuminate when Power
switch is placed in the On (I) position.
Monitor is not connected to an AC line
receptacle.
The AC power cord is defective.
The AC outlet is defective.
The power cord is not connected to the
monitor.
Blown fuses.
Connect the power cord to an AC line
receptacle.
Replace the power cord.
Use a different outlet.
Connect the power cord to the monitor.
Call Biomedical Engineering
Department.
Recorder does not function and the
Record indicator is off.
Recorder does not function and the
Record indicator flickers three short
flashes every five seconds.
Recorder functions however, Record
indicator flashes on and off every
second.
Recorder does not function; the Record
indicator is off; the message
PAPER
INCORRECTLY LOADED, RELOAD WITH
BLACK SQUARES DOWN
is shown in
maternal waveform area.
Recorder does not function and the
Record indicator is on.
Recorder is off, out of paper, or paper is
incorrectly loaded.
Recorder is in maternal-only mode.
Paper supply is low.
Paper loaded backwards.
Service required.
Press
Record
pushbutton; or install/re-
install paper (see
“Loading Strip Chart
Recorder Paper”
on page 5-2), then
press
Record
button.
Press
Record
pushbutton to turn on.
Install paper.
Re-install paper.
Call Biomedical Engineering
Department.
Incorrect time and date.
Time incorrectly set.
Clock circuit or battery fault.
Access the General Setup screen and
reset the time and date.
Call Biomedical Engineering
Department.
No heartbeat or pulse sounds.
Volume set too low.
Transducer not connected or is loose.
Press the Volume
pushbuttons or
access the respective setup screen(s)
(FECG, US, or US2) to increase the
volume.
Ensure that each transducer is securely
attached to monitor and applied to the
patient.
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......