6-6
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Functional Checkout Procedure: Connecting the Simulator
Connecting the Simulator
This part of the procedure prepares the simulator for use.
NOTE:
You must use a Model 325 Simulator for the functional checkout
procedure. (Monitors in the 120 Series do not work with Model 305
Simulators.)
1.
Ensure the Model 325
Power
switch is in the
off
position.
2.
Connect the Model 325 Simulator’s power cord to the power receptacle on the
rear panel of the simulator; plug the other end of the power cord into a properly
grounded wall outlet of appropriate voltage.
3.
Ensure the 120 Series Monitor
Power
switch is in the
off
position.
4.
Connect the simulator interconnect cable’s 50-pin end to the simulator’s
Fetal
Monitor
connector.
5.
Insert the green plug of the FECG/MECG adapter cable, cat. no. (REF)
1442AAO, into the monitor’s
FECG/MECG
connector.
6.
Connect the sub-cables of the other end of the simulator interconnect cable into
the color-coded connectors on the monitor/adapter:
ECG
,
US
, and
UA
.
7.
Turn
on
the Model 325 Simulator. Verify that the green
Power
indicator
illuminates.
8.
Turn
on
the 120 Series Monitor.
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......