General Maintenance: Maternal Blood Pressure Tests
Patient-to-Line Leakage for US
1.
Connect an US test body to the monitor’s front panel
US
input. The US test
body should be an ultrasound transducer wrapped in aluminum foil.
2.
Set the fixture’s
Leakage
switch to
Line
.
3.
Turn ON the monitor using the monitor’s power switch.
4.
Verify the following:
F
< 50
µ
A (100, 120, 220, 230, or 240 VAC monitors)
5.
Record the patient-to-ground leakage for the following conditions:
Table 9-12. Patient-to-Line Leakage Conditions for US
Neutral
Ground
Power (Polarity)
Closed
Closed
Normal
Closed
Closed
Reversed
Open
Closed
Normal
Open
Closed
Reversed