Revision B
120 Series Maternal/Fetal Monitor
5-7
2015590-001
Setup Procedures: User Setup Screens
IMPORTANT
EFFECTIVITY—All changes take effect immediately after a
selection is enacted in step six (6).
NOTE:
While any setup screen is displayed, the primary labor parameters remain
displayed. While using a setup screen, if an alarm condition exists for a
maternal parameter that is not displayed, only an audio alarm (if enabled) is
issued. Upon exiting the setup screen, the visual alarm indication is shown.
Example.
If the NBP Setup screen is displayed, the primary labor
parameters continue to be displayed as well as the maternal NBP area
of the screen. Under an alarm condition that affects NBP, both a visual
and audible (if enabled) alarm will be issued. Under an alarm
condition that affects MSpO
2
or MECG, only an audio alarm (if
enabled) is given while the NBP Setup screen remains displayed; once
the setup screen is exited, the visual alarm indication for MSpO
2
or
MECG will be shown.
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......