Revision B
120 Series Maternal/Fetal Monitor
3-11
2015590-001
Controls, Indicators, and Connectors: Front Panel Displays
MHR/P Area
The MHR/P area is summarized by
Figure 3-7
and
Table 3-7
.
Figure 3-7. MHR/P Display
Table 3-7. MHR/P Display
Name
Description
A
MHR/P Value
Up to three-digits indicate the MHR/P in beats per
minute.
B
MHR/P Alarm Setting
Indicator
This symbol provides information about the MHR/P
audio alarm and the MHR/P high/low alarm limit
settings.
: An alarm setting is turned off.
: All alarm settings are enabled.
C
MHR/P Mode Title
The mode title
MECG
indicates MECG is the MHR/
P source; the mode title
PULSE
indicates MSpO
2
or
NBP is used as the source. Select the mode title
softkey to access the MHR/P Setup screen.
D
Maternal Heartbeat
Indicator
Flashes with each detected valid heartbeat—for
MECG only.
89
MECG
C
D
B
A
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......