Revision B
120 Series Maternal/Fetal Monitor
4-55
2015590-001
Theory of Operation: UA/FECG Board
Table 4-21. Front End Motherboard Connector J1
Pin Number
Signal Name
Signal Description
1
No Connection
—
2
No Connection
—
3
AGND
Analog Ground
4
MECG1EN*
MECG Mode Enable
5
EXCITREF
+4 V Reference Ground
6
BRGEXCIT
+4 V Reference
7
PRESS–
Negative Input to Pressure Amplifier
8
STRAIN GAUGE*
IUP Mode Enable
9
AGND
Analog Ground
10
FRA
Fetal Right Arm Input
11
No Connection
—
12
B EN*
Mode Input (Unused)
13
FECG ENABLE*
FECG Mode Enable
14
MECG2 ENABLE*
Model 116 MECG Mode Enable
15
AGND
Analog Ground
16
PRESS+
Positive Input to Pressure Amplifier
17
TOCO*
TOCO Mode Enable
18
SENSE
Spare Analog Input
19
FLA
Fetal Left Arm Input
20
FRL
Fetal Right Leg Drive
Summary of Contents for Corometrics 126
Page 1: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 2: ......
Page 3: ...Corometrics 120 Series V3 5 SERVICE MANUAL MANUAL P N 2015590 001 REV B ...
Page 6: ...ii CE MARKING INFORMATION 0459 For Your Notes ...
Page 409: ...Revision B 120 Series Maternal Fetal Monitor C 1 2015590 001 Appendix C Drawings C ...
Page 410: ...C 2 120 Series Maternal Fetal Monitor Revision B 2015590 001 Drawings For your notes ...
Page 412: ......
Page 414: ......
Page 416: ......
Page 418: ......
Page 420: ......
Page 422: ......
Page 424: ......
Page 426: ......
Page 428: ......
Page 430: ......
Page 432: ......
Page 434: ......
Page 436: ......
Page 438: ......
Page 439: ......