11-14
120 Series Maternal/Fetal Monitor
Revision B
2015590-001
Peripheral Devices: Model 115-Compatible Communications Protocols
Table 11-3. Heart Rate Modes
Hr1 Mode
Hr2 Mode
Ascii Character
Hexadecimal Value
INOP
INOP
a
0
30
FECG
INOP
a
1
31
US
INOP
a
2
32
US
MECG
5
35
US
FECG
9
39
US
US2
;
3B
FECG
US
<
3C
INOP
MECG
=
3D
INOP
US2
>
3E
a
If HR2 mode is INOP and HR3 mode is MECG, then HR3 mode and data are sent as HR2 mode and data.
Table 11-1. Uterine Activity Modes
UA Mode
ASCII Character
Hexadecimal Value
INOP
0
30
TOCO
1
31
IUP
2
32