MFC
Operation
5.
129
Cl
2
mg/l, µg/l, ppb, ppm
Mem
mg/l, µg/l, ppb, ppm
mA/V
5 characters max. (any combination)
Selection of the displayed number format for mA/V
Sensor modules:
000.0 / 00.00 / 0000
Adjustment of the potential voltage at 3 electrode cells:
0–1000 mV
Selection of the µA signal measurement range for 3 electrode cells
and membrane sensors:
70 µA, 100 µA, 200 µA, 1000 µA
Setting the connected measurement signal:
0–20 mA, 4–20 mA, 0–10 V
Complete electrical connection according to this setting (see 9.
“Wiring Diagrams" on page 245)
Factor to adapt an external input signal:
0.1 to 4.0
only for sensor type Cl
2
++
:
Assignment of which pH module will be used to compensate for
the
2
measurement
pH (1) – pH (4) (equipped pH sensor modules only)
For sensor type CI-N combined only:
Assignment of which free CI
2
module will be used to compensate
for the combined CI
2
Cl
2
free (1) – Cl
2
free (4) (only equipped sensor modules
Unit
Cl2
Mem
mA/V
Format
mA/V
Upot
Cl2
µA measuring range
Cl2
Mem
Signal
mA/V
Factor
mA/V
pH Compensation
Cl2
Cl
2
free ref.
Mem
Summary of Contents for WALLACE & TIERNAN MFC
Page 1: ...WALLACE TIERNAN MULTI FUNCTION CONTROLLER MFC from version V 2 00 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 580 000 DE IM 0714 MFC Please note Original manual...
Page 6: ...6 WT 050 580 000 DE IM 0714 Introduction MFC 1...
Page 40: ...40 WT 050 580 000 DE IM 0714 Description MFC 3...
Page 247: ...MFC Certificates 10 247 10 Certificates...
Page 248: ...248 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 249: ...MFC Certificates 10 249...
Page 250: ...250 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 262: ...262 WT 050 580 000 DE IM 0714 Settings table MFC 11...