62
WT.050.580.000.DE.IM.0714
Functions
MFC
4.
Inserting into the flow-through
adapter
1
Insert the membrane sensor through the cover into the flow-
through adapter until it is resting mechanically on the inflow
mating connector and therefore receives a good flow. It may
be necessary to turn the flow-through adapter cover to set it in
the right position relative to the inflow mating connector.
Please note
Remove air bubbles from the membrane by lifting the membrane
sensor. They interfere with the measurement!
2
Connect the measuring signal cable to the measuring device.
Please note
If the measuring sensor is installed this way, it will function reliably
for approx. three to six months.
Storing the membrane sensor
1
Lift elastomer seal (H) and only then screw off the membrane
cap (E).
2
Wash the membrane cap (E) and electrode finger (D) with
clean (distilled) water.
3
Carefully dry the electrode finger (D) with absorbent paper.
4
Leave the membrane cap (E) to dry in a dust-free place.
5
Screw the dry membrane cap (E) loosely onto the electrode
shaft (B).
Please note
The membrane (G) may not touch the gold working electrode (J).
Restarting
See "Commissioning“.
Summary of Contents for WALLACE & TIERNAN MFC
Page 1: ...WALLACE TIERNAN MULTI FUNCTION CONTROLLER MFC from version V 2 00 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 580 000 DE IM 0714 MFC Please note Original manual...
Page 6: ...6 WT 050 580 000 DE IM 0714 Introduction MFC 1...
Page 40: ...40 WT 050 580 000 DE IM 0714 Description MFC 3...
Page 247: ...MFC Certificates 10 247 10 Certificates...
Page 248: ...248 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 249: ...MFC Certificates 10 249...
Page 250: ...250 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 262: ...262 WT 050 580 000 DE IM 0714 Settings table MFC 11...