MFC
Functions
4.
57
Installation notes
Ambient conditions
The following must be taken into account when installing the
membrane sensors:
•
Select the sample water extraction point that guarantees a
proper mixture of disinfectant and a bubble-free sample water
flow.
•
Keep the sample water extraction line as short as possible.
Please note
No water carrying lines made of copper piping may be installed.
These would distort the measurements.
•
The membrane sensors can only be used in the DEPOLOX
®
5
or VariaSens flow block assembly, which maintains the
constant sample water flow.
Because not all membrane sensors are designed for
pressurised operation, the operating conditions and the sensor
technical data must be reviewed and coordinated before the
sensors are installed in pressurised flow block assemblies!
•
If the membrane sensor is not installed right next to the MFC,
the measuring cell cable can be lengthened up to no more
than 50 m with a three-core, shielded cable. Ready- made
extension cables up to 15 m are available for this (see 8.
“Complete Devices, Retrofit Kits and Spare Parts" on
page 215).
Use terminal strips with right or left side connection routes
depending on the module slot in use. Find information on
connecting the sensor to the sensor module in the wiring diagrams
under 9. “Wiring Diagrams" on page 245.
Summary of Contents for WALLACE & TIERNAN MFC
Page 1: ...WALLACE TIERNAN MULTI FUNCTION CONTROLLER MFC from version V 2 00 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 580 000 DE IM 0714 MFC Please note Original manual...
Page 6: ...6 WT 050 580 000 DE IM 0714 Introduction MFC 1...
Page 40: ...40 WT 050 580 000 DE IM 0714 Description MFC 3...
Page 247: ...MFC Certificates 10 247 10 Certificates...
Page 248: ...248 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 249: ...MFC Certificates 10 249...
Page 250: ...250 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 262: ...262 WT 050 580 000 DE IM 0714 Settings table MFC 11...