Sensor measuring module
The sensor measuring module consists of:
•
Sensor (not in 3 electrode cells, mA/V input)
•
Sensor cable with watertight housing cable
duct (not in 3 electrode cells, mA/V input)
•
Pre-calibrated plug-in card
Due to the modular design, simple retrofitting and configuration of
sensor measuring modules in accordance with the plug-and-play
principle is possible at any time.
All sensor measuring modules and retrofit kits for Cl
2
, pH, mV, F
-
,
etc. can be plugged into module slots 1 to 4.
This configuration determines the MFC's functionality (see 4.2
“Measurement inputs" on page 46).
Please note
Module slot 5 is a control signal input with a MA/V sensor
measuring modules and is therefore not freely selectable.
A Sensor cable
B Housing duct
C Sensor
D Plug-in card
A
B
C
D