MFC
Operation
5.
159
5.4.5
ORP calibration (mV)
Please note
ORP sensors have long running-in times. After calibration with a
buffer solution, it can therefore take several hours until the value
has stabilised.
1
Starting from the basic display in the main menu select the
"Calibration" menu.
2
Select the menu item “mV (1) ... mV (4)“ for the measurement
you wish to perform.
The "Cal. mV" parameter is selected.
3
Place one of the supplied beakers into the bottom clip and fill
with the buffer solution or clamp a bag with buffer solution into
the bottom clip.
4
Pull or unscrew the ORP sensor from the lid of the cell body.
5
Dip the ORP sensor through the top clip at least 2 cm deep
into the buffer solution until the indicated pH value remains
constant.
6
Use the arrow keys to enter the determined value.
7
Save the value by pressing "Enter".
8
Remove the ORP sensor from the top clip.
9
Insert or screw the ORP sensor into the lid of the cell body.
ORP calibration is then concluded.
Summary of Contents for WALLACE & TIERNAN MFC
Page 1: ...WALLACE TIERNAN MULTI FUNCTION CONTROLLER MFC from version V 2 00 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 580 000 DE IM 0714 MFC Please note Original manual...
Page 6: ...6 WT 050 580 000 DE IM 0714 Introduction MFC 1...
Page 40: ...40 WT 050 580 000 DE IM 0714 Description MFC 3...
Page 247: ...MFC Certificates 10 247 10 Certificates...
Page 248: ...248 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 249: ...MFC Certificates 10 249...
Page 250: ...250 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 262: ...262 WT 050 580 000 DE IM 0714 Settings table MFC 11...