152
WT.050.580.000.DE.IM.0714
Operation
MFC
5.
Calibration aids
Two clips are installed in the housing cover. These clips can be
inserted into the rear panel of the housing.
The clip (A) for the sensor will be inserted into the upper catch.
When the electrodes are calibrated in the beakerwith the
calibration solution, the second clip (B) will be inserted into the
catch (left figure). The lower clip position is provided for calibrating
with the calibration solution bag (right figure).
A Upper clip
B Lower clip
Calibration configuration for the sensors that are calibrated outside
the cell body: pH, Redox, fluoride and conductivity
A Top clip with electrode mount
B Sensor
C Beaker
D Lower clip
E Ball valve on sample water inlet
(also on the sample water drain in the pressurised version)
A
B
B
A
A
B
C
D
E