204
WT.050.580.000.DE.IM.0714
Maintenance
MFC
7.
Checking for tightness (daily)
Check the entire measuring device including all screw connections
for leakage. Repair any leakage points immediately!
Please note
Air bubbles in the sample water influence the measuring accuracy.
The cause must be determined and remedied.
Checking the cell sand (weekly)
Check that there is sufficient sand in the cell body.
The cell sand must be swirled around in the bottom section of the
cell body.
The cell sand is necessary for cleaning the chlorine sensor
electrodes and must be replenished or replaced when required.
(Refer to 6.3.2 “Filling the cell sand (DEPOLOX® 5 only)" on
page 182 and “Replacing cell sand in DEPOLOX® 5 three-
electrode cell" on page 205).
Please note
When fresh sand is replenished, the electrode current may
increase slightly for approximately 3 hours. Do not calibrate during
this time.
You must calibrate each time the cell sand is replaced.
The calibration must be checked after one day.
mV measurement
Check Redox in buffer solution
every 4 to 6 weeks
Conductivity
Check conductivity
every 4 to 6 weeks
Task
Period/ Interval
page
Summary of Contents for WALLACE & TIERNAN MFC
Page 1: ...WALLACE TIERNAN MULTI FUNCTION CONTROLLER MFC from version V 2 00 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 580 000 DE IM 0714 MFC Please note Original manual...
Page 6: ...6 WT 050 580 000 DE IM 0714 Introduction MFC 1...
Page 40: ...40 WT 050 580 000 DE IM 0714 Description MFC 3...
Page 247: ...MFC Certificates 10 247 10 Certificates...
Page 248: ...248 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 249: ...MFC Certificates 10 249...
Page 250: ...250 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 262: ...262 WT 050 580 000 DE IM 0714 Settings table MFC 11...