142
WT.050.580.000.DE.IM.0714
Operation
MFC
5.
System - Menu 1.7
Basic display
Refer to main menu
Menu select
Display of all available menus
Menu system
Display of all available system settings
Common
Time (hh:mm)
Current time
Date (dd.mm.yy)
Current date
Measure Filter
1)
off / low / high
Hold function
Off / On
(see chapter 5.4
“Calibration" on page 151)
Language
German, English, French, Dutch
System name
12 characters, each with character set
A-Z and digits 1-9 including special
characters
Safety
Feed delay
2)
00:00 – 10:00 (03min : 00s)
Samp. water delay
3)
00:00 – 10:00 (01min : 00s)
(sample water delay)
System password
four-digit numeric code
(activate with Softkey "LOCK" in the "Menu Select" window)
Calibration password* four-digit numeric code
(activate with Softkey "LOCK" in the "Menu Select" window)
* only if system password is set
Trend Graph
Channel 1 to 4
Assignment of a measured value to the
line diagram. The selected measured
value is plotted in the trend graph
(can be traced back up to 7 days).
All the following measured values are
possible: module 1 - 4, flow rate Wq,
external setpoint/dosing factor,
temperature, control signals
Yout from module 1, Ypi from module 1 or
Yout from module 2.
If "Off" is selected, that
channel is not recorded.
Please note
If Ym is available, the Yout value for module 1 is displayed in Ym.
Summary of Contents for WALLACE & TIERNAN MFC
Page 1: ...WALLACE TIERNAN MULTI FUNCTION CONTROLLER MFC from version V 2 00 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 580 000 DE IM 0714 MFC Please note Original manual...
Page 6: ...6 WT 050 580 000 DE IM 0714 Introduction MFC 1...
Page 40: ...40 WT 050 580 000 DE IM 0714 Description MFC 3...
Page 247: ...MFC Certificates 10 247 10 Certificates...
Page 248: ...248 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 249: ...MFC Certificates 10 249...
Page 250: ...250 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 262: ...262 WT 050 580 000 DE IM 0714 Settings table MFC 11...