MFC
Functions
4.
53
Installation notes
Ambient conditions
The following must be taken into account when installing the
fluoride measurement:
•
Select the sample water extraction point that guarantees a
proper mixture of disinfectant and a bubble-free sample water
flow.
•
This does not require a certain flow rate.
•
The electrode must be immersed at least 2 cm deep into the
sample water.
Please note
No water carrying lines made of copper piping may be installed.
These would distort the measurements.
•
If the fluoride electrode is not installed right next to the MFC,
extension cables with plug connectors are available (see 8.
“Complete Devices, Retrofit Kits and Spare Parts" on
page 215).
•
If an extension cable is used (max. 50 m), a impedance
converter (see 8. “Complete Devices, Retrofit Kits and Spare
Parts" on page 215) must be used on the electrode in order to
guarantee a stable measuring signal.
•
Because the fluoride electrode has a connecting cable, the
impedance converter must be installed between the electrode
extension cable and the connecting cable.
Commissioning
Please note
The fluoride-sensitive membrane of the electrode is protected by a
rubber cap. The rubber cap must be removed before the electrode
is immersed in the sample water.
To prevent damages, do not touch the membrane.
Summary of Contents for WALLACE & TIERNAN MFC
Page 1: ...WALLACE TIERNAN MULTI FUNCTION CONTROLLER MFC from version V 2 00 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 580 000 DE IM 0714 MFC Please note Original manual...
Page 6: ...6 WT 050 580 000 DE IM 0714 Introduction MFC 1...
Page 40: ...40 WT 050 580 000 DE IM 0714 Description MFC 3...
Page 247: ...MFC Certificates 10 247 10 Certificates...
Page 248: ...248 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 249: ...MFC Certificates 10 249...
Page 250: ...250 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 262: ...262 WT 050 580 000 DE IM 0714 Settings table MFC 11...