52
WT.050.580.000.DE.IM.0714
Functions
MFC
4.
Installing the mV sensor
in the flow block assembly
1
Install the mV sensor in the corresponding opening on the flow
block assembly in theY flow-through adapter.
2
Screw the sensor cable marked "mV" into place.
3
The mV sensor can be calibrated for the first time after approx.
two to three hours running-in time.
Please note
The calibration must be checked after one day.
4.2.4
Fluoride Measurement
The fluoride measurement with fluoride ionic-sensitive electrodes
is used to continuously determine the fluorides in solutions. The
measurement medium's pH value must be between pH 4 and pH
8.5, in order to obtain a correct result.
Caution!
Quick and repetitive changes in temperature cause the potential to
change continuously, which can lead to electrode malfunction.
Please note
The reference system is filled with electrolyte. The integrated
storage tank must be replenished routinely with the solution.
The fluoride measuring system consists of:
•
Fluoride module with a solidly soldered connecting cable
•
Fluoride single-rod electrode
The fluoride module can be installed to slots 1 to 4 on the MFC.
Various controller functions are available depending on the slot
and application selected.
The electrodes can be installed on the DEPOLOX
®
5/
VariaSens flow block assembly or in separate Y flow-through
adapters.
Summary of Contents for WALLACE & TIERNAN MFC
Page 1: ...WALLACE TIERNAN MULTI FUNCTION CONTROLLER MFC from version V 2 00 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 580 000 DE IM 0714 MFC Please note Original manual...
Page 6: ...6 WT 050 580 000 DE IM 0714 Introduction MFC 1...
Page 40: ...40 WT 050 580 000 DE IM 0714 Description MFC 3...
Page 247: ...MFC Certificates 10 247 10 Certificates...
Page 248: ...248 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 249: ...MFC Certificates 10 249...
Page 250: ...250 WT 050 580 000 DE IM 0714 Certificates MFC 10...
Page 262: ...262 WT 050 580 000 DE IM 0714 Settings table MFC 11...