H Software history (NAMUR
recommendation NE53)
This appendix documents the change history of the Model 5700 transmitter software with configurable
outputs.
Document numbers refer to the English versions. Other languages have different part numbers but matching
revision letters.
September 2014, version 1.0
Document revision 20025166, Rev. AA - Initial release
February 2015, version 1.10
Document revision 20025166, Rev. AA
Software improvements:
•
Improved performance so the transmitter no longer spontaneously reboots after three weeks of
continuous operation.
•
Improved support for file transfer between the PC and the transmitter with ProLink III through 485
terminals.
•
Improved performance so that the progress bar no longer freezes when downloading a service file to a
USB drive.
•
Corrected the default value for the pressure compensation factor for density for the CMF350 sensor to -9.0
E
6
g/cm
3
/psi (-0.0000090).
September 2015, version 1.20
Document revision 20025166, Rev. AA
Software improvements:
•
Corrected flow rate report when LD optimization is enabled.
•
Added ability to configure the display precision to 0 (number of decimal places).
Feature additions:
•
Added support for the F100P sensor
•
Added support for OIML-compliant display for future OIML-approved transmitter
•
Added Modbus registers to match the defaults for integration with the microLoad.net preset controller
(FMC Technologies)
•
Similar functionality to ETO 23182 for Model 2700 transmitters
•
Improved Russian and Chinese translations
April 2016, version 1.30
Document revision 20025166, Rev. AA
Software improvements:
Configuration and Use Manual
Software history (NAMUR recommendation NE53)
MMI-20025166
March 2019
Configuration and Use Manual
351