5. Optional: Pause the batch at any time.
•
Using the display: Choose
Menu
>
Batcher
>
Run Batch
>
Batch Control
and select
Pause
.
•
Using ProLink III: Choose
Device Tools
>
Application Control
>
Run Batcher
and click
Pause
Batcher
.
Pausing a batch allows you to resume it later.
6. Optional: Resume the batch after it has been paused.
•
Using the display: Choose
Menu
>
Batcher
>
Run Batch
>
Batch Control
and select
Resume
.
•
Using ProLink III: Choose
Device Tools
>
Application Control
>
Run Batcher
and click
Resume
Batch
.
7. Optional: End the batch manually at any time before it reaches the target.
•
Using the display: Choose
Menu
>
Batcher
>
Run Batch
>
Batch Control
and select
End
.
•
Using ProLink III: Choose
Device Tools
>
Application Control
>
Run Batcher
and click
End Batch
.
If you end a batch manually, you will not be able to resume it. You can end the batch while it is running
and while it is paused.
8. If you did not end the batch manually, wait until the batch ends automatically.
•
Using the display: Choose
Menu
>
Batcher
>
Run Batch
>
Batch Control
.
•
Using ProLink III: Choose
Device Tools
>
Application Control
>
Run Batcher
.
The batch ends automatically when it reaches the target or when it reaches the configured maximum
fill time.
9. Optional: Inhibit batch.
•
This function is not available from the display.
•
Using ProLink III: Choose
Device Tools
>
Application Control
>
Run Batcher
.
While the
Inhibit Batch
input is active, a batch will not be allowed to start. Attempting to start a batch
while the batch is inhibited will cause the Start Not Okay alarm to be posted.
10. Optional: Inhibit totalizer.
•
This function is not available from the display.
•
Using ProLink III: Choose
Device Tools
>
Application Control
>
Run Batcher
.
When the
Inhibit Totalizer
input is active, the
Batch Total
will not increment. A timeout alarm will
occur if the
Batch Total
is inhibited beyond the max batch time. The batch inventory for the current
preset will not increment while
Inhibit Totalizer
is active.
11. Optional: Inhibit flow.
•
This function is not available from the display.
•
Using ProLink III: Choose
Device Tools
>
Application Control
>
Run Batcher
.
When the
Inhibit Flow
input is active, the batch will ignore the flow rate of the assigned flow source,
allowing a batch to end. The batch total will not increment when flow is inhibited.
Operation using the batcher
Configuration and Use Manual
March 2019
MMI-20025166
196