17.13.54 mA Output 3 Fixed
Alert
Output Fixed
Cause
The output is configured to send a constant value. A loop test may be in progress.
Recommended actions
1. Check to see if a loop test is in process (the output is fixed).
2. Exit mA output trim, if applicable.
17.13.55 mA Output 3 Saturated
Alert
Output Saturated
Cause
The calculated output value is outside the range of the output.
Recommended actions
1. Check the settings of
Upper Range Value
and
Lower Range Value
.
Configure Lower Range Value (LRV) and Upper Range Value (URV) for the mA Output
2. Check your process conditions against the values reported by the device.
3. Ensure that both devices are using the same measurement unit.
4. Purge the sensor tubes.
17.13.56 Moderate Two Phase Flow
Alert
Process Aberration
Cause
The transmitter has detected moderate two-phase flow.
Recommended actions
Check your process conditions against the values reported by the device.
17.13.57 No Permanent License
Alert
Data Loss Possible
Troubleshooting
Configuration and Use Manual
March 2019
MMI-20025166
280
Micro Motion Model 5700 Transmitters with Configurable Outputs