Figure 8-1: Effect of mA Output Direction on an mA Output: Lower Range Value = 0
mA Output Direction = Normal
mA output
-x
0
x
Reverse flow
Forward flow
20
12
4
mA Output Direction = Absolute Value
mA output
-x
0
x
Reverse flow
Forward flow
20
12
4
•
Lower Range Value
= 0
•
Upper Range Value
= x
Figure 8-2: Effect of mA Output Direction on an mA Output: Lower Range Value < 0
mA Output Direction = Normal
mA output
-x
0
x
Reverse flow
Forward flow
20
12
4
mA Output Direction = Absolute Value
mA output
-x
0
x
Reverse flow
Forward flow
20
12
4
•
Lower Range Value
=
−
x
•
Upper Range Value
= x
mA Output Direction
= Normal and
Lower Range Value
= 0
Configuration:
Integrate the meter with the control system
Configuration and Use Manual
March 2019
MMI-20025166
142