Severity configurable: Yes
Fault Timeout applicable: Yes
Table 7-8: Extreme Primary Purpose Variable conditions
Name
Ignorable
[005] Mass Flow Rate Overrange
No
[008] Density Overrange
No
Function Check Failed or Smart Meter Verification Aborted
Default severity: Maintenance Required
Severity configurable: Yes
Fault Timeout applicable: No
Table 7-9: Function Check Failed or SMV Aborted conditions
Name
Ignorable
[010] Calibration Failed
No
[034] Smart Meter Verification Failed
Yes
[035] Smart Meter Verification Aborted
Yes
Function Check in Progress
Default severity: Function Check
Severity configurable: No
Fault Timeout applicable: No
Table 7-10: Function Check in Progress conditions
Name
Ignorable
[104] Calibration in Progress
No
[131] Smart Meter Verification in Progress
Yes
Output Fixed
Default severity: Function Check
Severity configurable: Yes
Fault Timeout applicable: No
Table 7-11: Output Fixed conditions
Name
Ignorable
[101] mA Output 1 Fixed
Yes
[111] Frequency Output 1 Fixed
No
[114] mA Output 2 Fixed
No
[118] Discrete Output 1 Fixed
No
[119] Discrete Output 2 Fixed
No
Configuration and Use Manual
Configure device options and preferences
MMI-20025166
March 2019
Configuration and Use Manual
131