Note
Unlike earlier transmitters, the 5700 derives
Sensor Type
from the user-specified values for FCF and K1
in combination with an internal ID.
2. Set the flow calibration factor:
FCF
(also called
Flow Cal
or
Flow Calibration Factor
). Be sure to include
all decimal points.
3. Set the density characterization parameters:
D1
,
D2
,
TC
,
K1
,
K2
, and
FD
. (
TC
is sometimes shown as
DT
.)
4. Apply the changes as required by the tool you are using.
The transmitter identifies your sensor type, and characterization parameters are adjusted as required:
•
If
Sensor Type
changed from Curved Tube to Straight Tube, five characterization parameters are
added to the list.
•
If
Sensor Type
changed from Straight Tube to Curved Tube, five characterization parameters are
removed from the list.
•
If
Sensor Type
did not change, the list of characterization parameters does not change.
5. T-Series sensors only: Set the additional characterization parameters listed below.
Characterization parameter type
Parameters
Flow
FTG
,
FFQ
Density
DTG
,
DFQ1
,
DFQ2
2.9.1 Sample sensor tags
Figure 2-1: Tag on newer curved-tube sensors (all sensors except T-Series)
Quick start
Configuration and Use Manual
March 2019
MMI-20025166
14
Micro Motion Model 5700 Transmitters with Configurable Outputs