Operating Manual 

Pressure Transmitter for Shipbuilding and Offshore  
Applications 

DMK 457, DMK 458 and DMP 457 

Headquarter Western Europe / International 

BD SENSORS GmbH 
BD-Sensors-Str. 1 
D - 95199 Thierstein 
Germany 

Tel:  +49 (0) 92 35 / 98 11-0 
Fax: +49 (0) 92 35 / 98 11-11 

Headquarter Eastern 
Europe 

BD SENSORS s.r.o. 
Hradištská 817 
CZ - 687 08 Buchlovice 
Czech Republic 

Tel:  +42 (0) 5 72 / 4 11-0 11
Fax: +42 (0) 5 72 / 4 11-4 97

Russia 

BD SENSORS RUS 
39a, Varshavskoe shosse 
RU - Moscow 117105 
Russia 

Tel:  +7 (0) 9 59 81 / 09 63 
Fax: +7 (0) 9 57 95 / 07 21 

further agencies in: 

EUROPE

 

  Belgium 

  Denmark 

  France 

  Greece 

  Italy 

  Lithuania 

  Luxemburg 

  Netherlands 

  Norway 

 

  Poland 

  Romania 

  Sweden 

  Switzerland 

  Slovakia 

  Spain 

  Turkey 

  UK 

  Ukraine 

ASIA 

  Iran 

  Israel 

  Kazakhstan 

  Taiwan 

  Thailand 

AUSTRALIA 

The  addresses  of  our  distribution  partners  are  listed  on  our 
homepage 

www.bdsensors.com

. It is possible to download 

data  sheets,  operating  manuals,  ordering  codes  and  certifi-
cates, as well. 

1. General information 

1.1 Information on the operating manual 

This  operating  manual  contains  important  information  on 
proper  usage  of  the  device.  Read  this  operating  manual 
carefully  before  installing  and  starting  up  the  pressure 
measuring device. 

Adhere to the safety  notes and  operating instructions which 
are  given  in  the  operating  manual.  Additionally  applicable 
regulations  regarding  occupational  safety,  accident  preven-
tion  as  well  as  national  installation  standards  and  engineer-
ing rules must be complied with! 

This  operating  manual  is  part  of  the  device,  must  be  kept 
nearest its location, always accessible to all employees. 

This  operating  manual  is  copyrighted.  The  contents  of  this 
operating manual reflect the version available at the time of 
printing. It has been issued to our best knowledge. However, 
errors  may  have  occurred.  BD  SENSORS  is  not  liable  for 
any incorrect statements and their effects. 

– Technical modifications reserved – 

1.2 Symbols used 

U

U

U

U

 

DANGER!  –  dangerous  situation,  which  may  result  in 
death or serious injuries 

U

U

U

U

 

WARNING!  –  potentially  dangerous  situation,  which 
may result in death or serious injuries 

U

U

U

U

 

CAUTION! – potentially dangerous situation, which may 
result in minor injuries 

 

!

 

CAUTION! – potentially dangerous situation, which may 
result in physical damage  

 

NOTE  –  tips  and  information  to  ensure  a  failure-free 
operation 

1.3 Target group 

U

U

U

U

 

WARNING! To avoid operator hazards and damages of 
the device, the following instructions have to be worked 
out by qualified technical personnel. 

1.4 Limitation of liability 

By  non-observance  of  the  operating  manual,  inappropriate 
use,  modification  or  damage,  no  liability  is  assumed  and 
warranty claims will be excluded. 

1.5 Intended use 

-  Pressure  transmitters  DMK 456,  DMK 457,  DMK 458 

and  DMP 457  have  been  designed  for  hard  conditions 
especially  in  shipbuilding  and  offshore  applications. 
They  are  suitable  for  measuring  tasks  with  fluids  and 
gases.  Typical  applications  of  DMK 456  and  DMK 458 
are pressure monitoring for loading and discharge proc-
esses  as  well  as  level  measurement  for  ballast  and 
product  storage  tanks.  Preferred  areas  of  usage  for 
DMK 457  are  gears,  compressors,  boilers,  pneumatic 
controls,  elevators  and  oxygen  applications.  With  me-
chanical  versions  G1/2"  open  port  or  G1/2"  flush 
DIN 3852 the DMK 457 is especially suited for  viscous, 
pasty  or  contaminated  media  due  to  the  easily  reach-
able  ceramic  diaphragm.  Preferred  areas  of  usage  for 
DMP 457  are  diesel  engines,  gears,  compressors, 
pumps,  boilers,  hydraulic  and  pneumatic  controls  as 
well  as  elevators.  The  pressure  transmitters  DMK 456, 
DMK 457,  DMK 458  and  DMP 457  fulfil  the  require-
ments  of  Germanischer  Lloyd  (GL)  as  standard. 
DMK 457 as well as DMP 457 is additionally certificated 
by Det Norske Veritas (DNV). The certificates are avail-
able 

for 

download 

on 

our 

homepage: 

http:// 

www.bdsensors.com/proucts/download/certificates 

-  It is the operator's responsibility to check and verify the 

suitability  of  the  device  for  the  intended  application.  If 
any doubts remain, please contact our sales department 
in  order  to  ensure  proper  usage.  BD  SENSORS  is  not 
liable for any incorrect selections and their effects! 

-  Permissible media are gases or liquids, specified in the 

data  sheet.  In  addition  it  has  to  be  ensured,  that  this 
medium is compatible with the media wetted parts.  

-  The  technical  data  listed  in  the  current  data  sheet  are 

engaging and must be complied with. If the data sheet is 
not  available,  please  order  or  download  it  from  our 
homepage:  http://www.bdsensors.com/products/down-
load/datasheets 

U

U

U

U

 

WARNING!

 

Danger through improper usage!  

1.6 Package contents 

Please verify that all listed parts are undamaged included in 
the  delivery  and  check  for  consistency  specified  in  your 
order: 

-  pressure transmitter  

-  mounting instructions 

-  with option SIL2 version:  

Functional Safety Manual, Functional Safety Data 
Sheet

, SIL Declaration of Conformity 

2. Product identification 

The  device  can  be  identified  by  its  manufacturing  label.  It 
provides  the  most  important  data.  By  the  ordering  code  the 
product can be clearly identified.

 

Fig. 1 manufacturing label 

!

 

The manufacturing label must not be removed from the 
device! 

3. Mechanical installation 

3.1 Mounting and safety instructions 

U

U

U

U

 

WARNING!  Install  the  device  only  when  depressurized 
and currentless! 

U

U

U

U

 

WARNING!  This  device  may  only  be  installed  by 
qualified  technical  personnel  who  has  read  and  under-
stood the operating manual! 

U

U

U

U

 

DANGER!  Explosion  hazard,  with  devices  for  oxygen 
applications, when used improperly. To ensure a usage 
without danger, the following points must be adhered to: 

-  Make sure, your device has been ordered and de-

livered  as  a  special  version  for  oxygen  applica-
tions. You can check the manufacturing label (see 
figure 1). If the ordering code ends with "007", then 
the device is suitable for oxygen applications. 

-  At time of delivery the device is packed into a plas-

tic bag in order to  prevent it from impurity. Please 
observe  the  indication  label  "Device  for  oxygen, 
unpack  only  directly  before  assembling".  Also, 
avoid  any  skin  contacts  during  unpacking  and  as-
sembly, in order to prevent greasy residues on the 
device. 

-  During installation, the respective explosion protec-

tion  regulations  have  to  be  met.  Check,  if  ATEX-
approval  is  necessary  for  this  type  (oxygen)  de-
vice. (the delivered device has no ATEX-approval) 

-  Note  the  entire  design  requirements  meet  the 

standard demand of BAM (DIN 19247). 

-  For devices with oxygen capability up to 50 bar, o-

rings  V747-75  with  BAM-approval  are  being  used. 
The  max.  capabilities  allowed,  are  40 bar/130° C 
and 50 bar/100° C. 

-  For devices with oxygen capability over 50 bar, o-

rings  FKM  90  are  being  used.  These  have  been 
tested up to 95° C and 215 bar in the scientific coal 
research institute in Ostrava – CZ. 

!

 

Handle 

this 

high-sensitive 

electronic 

precision 

measuring  device  with  care,  both  in  packed  and 
unpacked condition! 

!

 

There are no modifications/changes to be made on the 
device.  

!

 

Do not throw the package/device! 

!

 

To  avoid  damaging  the  diaphragm,  remove  packaging 
and  protective  cap  directly  before  starting  assembly. 
The delivered protective cap has to be stored! 

!

 

Place  the  protective  cap  on  the  pressure  port  again 
immediately after disassembling. 

!

 

Handle  the  unprotected  diaphragm  very  carefully  - it is 
very sensitive and may be easily damaged. 

!

 

Do  not  use  any  force  when  installing  the  device  to 
prevent damage of the device and the plant! 

!

 

For  installations  outdoor  or  in  moist  surroundings,  the 
following points have to be observed: 

-  To prevent moisture  admission in the plug the de-

vice should be installed electrically after mounting, 
at once. Otherwise a moisture admission has to be 
blocked  e.g.  by  using  a  suitable  protection  cap. 
(The  ingress  protection  in  the  data  sheet  is  valid 
for the connected device.)  

-  Choose  an  assembly  position,  which  allows  the 

flow-off of splashed water and condensation. Avoid 
permanent fluid at sealing surfaces! 

-  When  using  a  cable  gland  device,  turn  the  outgo-

ing cable downwards. If the cable has to be turned 
upwards,  then  point  it  downward  so  the  moisture 
can drain. 

-  Install the device in such a way that it is protected 

from direct solar irradiation. Direct solar irradiation 
can  lead  to  the  permissible  operating  temperature 
being overstepped in the worst case. Through this, 
the  operability  of  the  device  can  be  affected  or 
damaged. If the internal pressure increases due to 
solar  irradiation,  temporary  measurement  errors 
may occur. 

!

 

For devices with gauge reference in the housing (small 
hole next to the electrical connection), install the device 
in  such  a  way,  that  the  gauge  reference  is  protected 
from dirt and moisture. Should the device be exposed to 
fluid  admission,  the  functionality  will  be  blocked  by  the 
gauge  reference.  An  exact  measurement  in  this  condi-
tion  is  not  possible.  Furthermore  this  can  lead  to  dam-
ages on the device. 

 

Take  note  that  no  assembly  stress  occurs  at  the 
pressure  port,  since  this  may  cause  a  shifting  of  the 
characteristic curve. This is especially important for very 
small  pressure  ranges  as  well  as  for  devices  with  a 
pressure port made of plastic. 

 

In hydraulic systems, position the device in such a way 
that the pressure port points upward (ventilation). 

 

Provide  a  cooling  line  when  using  the  device  in  steam 
piping. 

 

If  there  is  any  danger  of  damage  by  lightning  or 
overpressure  when  the  device  is  installed  outdoor,  we 
suggest putting a sufficiently dimensioned overpressure 
protection between the supply or switch cabinet and the 
device. 

3.2 General installation steps 

-  Carefully  remove  the  pressure  measuring  device  from 

the package and dispose of the package properly. 

-  Go ahead as detailed in the specific instructions below.  

 

ordering code 

connector pinout 

type designation 

supply 

nominal 

pressure 

range 

signal 

serial 

number 

 